Description
ADWA replacement probe for AD31 EC meter – replace the probe, keep your meter
The AD32P lets you replace the probe on your ADWA AD31 conductivity meter without replacing the whole instrument. This original spare part also fits the AD32 and suits growers who use their existing meter to check water and nutrient solutions.
Conductivity and temperature from one probe
The probe measures electrical conductivity and incorporates a temperature sensor. The connected meter uses the temperature data for automatic temperature compensation. It also calculates a TDS value from conductivity as a measure of dissolved solids. The AD32P is not a pH electrode and requires a compatible meter to operate.
Replace the probe, then recalibrate
Undo the threaded probe housing, fit the replacement probe and carefully tighten the housing again. Correct assembly is important to maintain the meter’s watertight seal.
After replacing the probe, calibrate your AD31 using a 1413 µS/cm EC calibration solution. This is equivalent to 1.413 mS/cm. pH buffer solutions are not suitable for this process. When fitted to an AD32, the required EC calibration value is 12.88 mS/cm instead.
Technical details:
- Product type: Replaceable EC probe with temperature measurement
- Model: ADWA AD32P
- Compatibility: ADWA AD31 and AD32
- Temperature sensor: Integrated
- EC calibration value for AD31: 1413 µS/cm, equivalent to 1.413 mS/cm
- EC calibration value for AD32: 12.88 mS/cm
- Package contents: 1 × ADWA AD32P replacement probe
Sold and shipped by Growmart (Hamburg, Germany), which delivers across Europe with DHL. Prices are set by the vendor and you complete your purchase on their site.